Released on 2026-11-26. Mostly waveforms a second: the screen follows the panel's refresh, the drawing of Normal,
Peak, High res, Average and zoom runs on the four drawing cores, the short time bases interpolate in fixed point, and
the run state box shows the waveforms a second. Several of these ideas come from fumingcaps' fork (0.2.1), named at
each entry; its timing shortcuts that change the picture or the measured values were not taken. On a DHO924S, measured
the same day with tools/device-bench.py, this build reaches the screen with more waveforms a second than that fork
at 14 of 18 views, and fewer at 4 channels at 5 ns/div and in Peak detect. Further: the Bode page (number pad, Lin /
Log sweep, CSV, Vrms), the rear USB port as a network link (off by default), fixes from the forum (the trigger on the
other slope of a slow, noisy signal, #578 and #581; persistence, the picture while the scope waits) and from W4R's OOS
Plus fork. All of it is tested on the host; marked "(unit)" where it also ran on a DHO924S. A build of main from the
same day ran tools/device-test.py on a DHO924S.

Known: after a STOP with a deep memory (10 M to 125 M points), :WAVeform:MODE RAW sometimes answers the 3600 points
of the screen instead of the raw record around it; the screen and the other modes are not affected. Seen on a DHO924S
in tools/device-test.py (raw_window_is_one_unbroken_signal), not every time; the cause is still open.

Added:

  • Bode: a tap on the Amplitude opens the number pad, so the generator takes any amplitude from 10 mVpp to 20 Vpp
    (350 mVpp, 1.25 Vpp), not only the 1-2-5 steps of - and +; from such a value - and + go to the next step
    (#52, Measurements and Analysers).
  • The Bode plot shows the values of its gain and phase axes at round steps (#387,
    Measurements and Analysers).
  • Bode: with the menu closed, the Multipurpose 2 knob moves the cursor over the plot; before, only the menu's Cursor
    row did, and the menu covered half of the plot
    (#50, Measurements and Analysers).
  • Bode: Sweep Lin / Log now spaces the frequencies of every sweep and sets the frequency axis to match. Gain
    unit
    shows the gain in dB or in V/V
    (Measurements and Analysers).
  • Bode: Save CSV writes the points of the last sweep to the storage: frequency, gain and phase, one line a point
    (#53, Measurements and Analysers).
  • The Manchester decoder shows the frame check of an Ethernet frame (10BASE-T): the last four bytes of the frame as a
    check, marked where the CRC-32 over the frame is another. As bytes alone a good frame and one with a byte changed
    looked alike. A frame the record cuts off, or one shorter than a frame can be, has no check.
  • A tap on a value in the measurement strip opens the list of selected measurements with that one in focus, to remove
    it or change its settings (Measurements).
  • The rear USB port can be a network link to one computer: System > Network > Remote services > USB network (off
    by default; tried on one scope with a Mac and Windows 11, not yet with Linux). The computer sees a network adapter,
    the scope is at 192.168.235.1 and every remote service and SSH work over the cable
    (Remote control, #235) (unit).
  • While the scope runs, the run state box at the top left shows the waveforms a second that reach the screen and the
    share of the time they capture, for example 14.8k wfm/s 0.7% (First Steps). The idea
    comes from fumingcaps' fork.

Removed:

  • The image's busybox no longer holds telnetd, the telnet client, httpd or tc, and nc no longer listens
    (nc -l); nothing in OOS started them. busybox is now built from its source with a configuration of OOS
    (network.md, ADR 0078, #331).
  • System > Opt. memory and Probe optional memory: hardware segments above 256 MiB read back the samples of the
    segment 256 MiB below them, so the upper half is not usable. Hardware segments use the lower 256 MiB again, as
    before 0.1.15
    (horizontal-config.md 1.8).
    A mem.upper line left in /oos/system.conf is kept and does nothing (unit).

Changed:

  • oos-ui is built for the scope's small cores (Cortex-A53), where nearly every frame is drawn: 5 to 10 % more
    waveforms a second (DHO924S, 1 channel at 50 ns/div 7,550 -> 8,280), 32 instead of 28 pictures a second with 4
    channels at 5 ns/div. The idea comes from fumingcaps' fork.
  • While the scope runs, the acquisition thread runs at real-time priority and the fast cores and the GPU do not drop
    to their lowest clocks; STOP gives the clocks back. 2 to 7 % more waveforms a second at the short time bases
    (DHO924S, 1 channel at 5 ns/div 7,230 -> 7,710). perf.fifo = off and perf.clocks = off in system.conf turn
    them off (performance plan).
  • Average folds many more waveforms a second into its average, made on the four drawing cores, and the picture shows
    the one averaged trace instead of each step of the average one over the other; the picture rate holds with several
    channels on (Features).
  • Average folds its frames on the four drawing cores in runs, applied to the average in order: 2,240 -> 3,320
    waveforms a second with 4 channels at 1 us/div on a DHO924S, the same average up to rounding. The idea comes from
    fumingcaps' fork.
  • Batch mode: at short time bases the scope arms only as many frames as the screen's workers can draw in one picture
    and waits for them, instead of reading batches it then drops; acquired and dropped of :SYSTem:STATe? fall
    towards used (SCPI). 11 to 20 % more waveforms a second reach the screen (DHO924S, 1 channel at
    5 ns/div 7,230 -> 8,030; 4 channels at 200 ns/div 1,390 -> 1,670).
  • Acquisition: a batch's register writes wait once after each group instead of after every write (75 waits of at
    least 10 us a batch became 2); 8 to 13 % more waveforms a second reach the screen at the short time bases
    (DHO924S, 1 channel at 5 ns/div 6,060 -> 6,830).
  • Bode: - and + step the generator amplitude by 10 mV up to 100 mV and by 50 mV above, and Amp unit shows and takes
    it in Vpp or Vrms; :BODE:AMPLitude sets it in Vpp
    (Measurements and Analysers).
  • The short time bases (fewer samples than screen columns) interpolate their points in fixed point, as fumingcaps'
    fork does: about half the cost of a frame's points on the scope's small cores, more waveforms a second at 5 to
    200 ns/div. The trace stays within a pixel of the one before.
  • The screen is drawn at every third refresh of the panel (32.5 pictures a second on a DHO924S, whose panel runs at 97.4
    Hz) instead of on a 30 Hz timer of its own, so its pictures no longer beat against the panel; pacing.clock = timer
    in /oos/system.conf goes back to the timer, and a monitor on HDMI keeps it (ADR
    0010).
  • The GPU helper may run on every core instead of only the two fast ones it shared with the screen loop and the
    acquisition: in Average with 4 channels the screen loop no longer waits for it, about 10 % more waveforms a second.
  • In the menus a setting that is only off or on is one switch instead of the two buttons Off and On: a tap anywhere on
    it switches it (#55, First Steps).
  • Display (lines, not dots): a trace drawn from a record (short time bases, Average, Peak, High res, zoom) is as bright
    on its vertical edges as on its flat parts; it was up to four times dimmer on the edges. The pixels lit are the same.
  • The number pad, the generator's frequency pad and the hex pad of a bus trigger open large in the middle of the
    screen, with the value in a field above the keys; the cursor menu opens at the top, so its readouts stay in sight
    (First Steps).
  • Peak and High res draw their records on the four drawing cores, as Normal does, instead of one by one in the
    picture loop: more waveforms a second and a steadier picture rate with several channels on
    (Features) (unit).
  • An on/off switch that is on fills its whole field white with dark text, like a chosen option, and its slide switch
    turns dark with a white knob, so the state shows at a glance (First Steps).
  • The LAN and WLAN boxes of the top bar no longer show the scope's address by default: they show their name and a
    green dot while connected, or a dash without a link, and are only as wide as that, which leaves the top bar more
    room. A picture or photo of the screen no longer passes the address on. System > Network > Show address shows
    it again (Features).
  • The top bar holds Display and Storage; Measure and Cursor are their front-panel keys and entries of the start menu.
    The Measure key held adds the next usual measurement, as the press-and-hold on the Measure button did; that short
    menu is gone (Measurements and Analysers).
  • Display: at the short time bases, where a frame is drawn from its 3600 interpolated points, 11 to 16 % more
    waveforms a second reach the screen (DHO924S, 1 channel at 5 ns/div 6,060 -> 6,900; 4 channels at 200 ns/div
    1,150 -> 1,340); the picture is the same.
  • A trigger type keeps its own values (width, pattern, times, lower level) and gets them back when chosen again; each
    channel keeps its own trigger level as the source (First Steps).
  • With the zoom on, the main and the zoom picture are drawn on the four drawing cores, as without it, instead of one
    record at a time in the picture loop: more waveforms a second and a steadier picture rate
    (Features) (unit).

Fixed:

  • A channel switched off with its menu's On, not with its key, gives the vertical knobs to a channel still on, as the
    key does; before, V/div and the offset turned the channel that was off (#523,
    First Steps).
  • :CHANnel<n>:OFFSet and :SCALe keep the offset within what the scale can reach at once: a query in the same line
    answers the offset in force, and a line of commands gives the same offset whether or not a frame ran between them.
    Found by W4R's OOS Plus fork (SCPI and VXI-11).
  • Cursor: - and + on a voltage cursor (A and B in Volt, Y1 and Y2 in XY) move it up and down the right way round:
    • raises its value, as it moves a time cursor to the right (#248,
      First Steps).
  • The Cycle measurements (Cycle mean, RMS and area, Cyc top, base, amplitude, pk-pk, max and min) have a value when
    the one whole cycle on screen lies between two falling edges, as the period and the duty cycle already had; found
    by W4R's OOS Plus fork (Measurements and Analysers).
  • The decoders and the protocol trigger read a record taken in Peak detect. Its points are minimum and maximum in
    turn, so every edge passed the threshold three times: a LIN frame at 2.5 MSa/s came out with no sync measured and
    every byte wrong, while the same line decoded in Normal. They now read the mean of every point and its neighbour
    (Features).
  • The eye page no longer takes all the memory of the scope when a bit rate from a faster time base is kept at a slow
    one: a rate the record has under two samples a bit for is found again from the record when it was found from one,
    and a rate that was set stays and the page says that it needs a faster time base. The clock recovery takes no
    bit shorter than half a sample, which bounds the eye, the jitter figures and the TIE trend for every caller. Seen
    on a unit: the program was stopped by the kernel at 3.7 GB
    (Measurements and Analysers).
  • Display: at 10 us/div to 1 ms/div the screen again keeps its full picture rate on the panel's clock (it had dropped
    to 23 to 29 pictures a second since the frame loop follows every third vblank).
  • The history keys of the front panel step the right way: < goes back to older records, > forward to newer ones
    (front panel) (unit).
  • The Manchester decoder reads a long packet whose sender's clock is a little off the rate set. Its bit clock now
    follows the edge in the middle of every bit, not the first edge of the packet: a packet of 1500 bytes at 200 ppm from
    the rate went wrong from its byte 234 before, as a frame of 10BASE-T (880 bits) did on a unit with a wrong byte and
    a wrong frame check. A rate set 8 % off still reads the line.
  • While the scope ran with a record of more than 262144 points that came faster than it was measured, Frequency,
    Period, the widths, rise and fall time, Top and Base stayed on Computing; they now show the value of the last
    record that was measured (Measurements and Analysers).
  • The MIL-STD-1553 decoder reads a bus a few percent off its rate: it follows the edge in the middle of every bit
    from bit to bit. It sampled the 17 bits of a word on a 1 us grid from the sync, and on a bus 4 % slow the later
    bits fell into the wrong half: on a capture of a real bus 37 of 71 words were misread or had a parity error, now 2
    (the capture's own faults) (Features).
  • After a STOP of a deep record, :WAVeform:MODE NORMal returns what the screen shows also when a measurement needs
    samples, not only the window read at the STOP (SCPI) (unit).
  • With a display persistence time (100 ms, 500 ms and longer) the trace no longer flickers or pulses at slow time
    bases and deep memory. The picture faded on every screen refresh, also while it waited for the next record, so the
    newest trace faded out before the next one came. It now fades when a new record arrives, by the time since the last
    one: the newest trace stays at full brightness until the next record replaces it
    (First Steps).
  • Touch: two fingers held still on the plot are a pinch and no longer open the context menu of a long press (#590).
  • A remote command over SCPI (port 5555), VXI-11 or HiSLIP is answered as soon as it arrives, not at the next picture:
    *OPC? in Run takes a few milliseconds instead of 12 to 30 (SCPI and VXI-11) (unit).
  • In roll mode the state box at the top left says ROLL instead of TRIG'D, as does the line of the HDMI workspace
    (Features).
  • SCPI: *RST also sets the waveform transfer back (:WAV:SOURce, :MODE, :FORMat, :STARt, :STOP), so a
    page left by an earlier read no longer cuts the next one (docs/scpi.md).
  • A remote command that exists but gets a value it does not take (:TRIGger:HOLDoff 11, :CHANnel1:PROBe 0,
    :ACQuire:MDEPth 7, :TIMebase:SCALe x, ...) leaves -224,"Illegal parameter value" with the value in
    :SYSTem:ERRor?, no longer -113,"Undefined header", which now only means the command is not known
    (SCPI and VXI-11).
  • A line of a setup file that is reported as bad leaves the setup as it was: before, the Bode plot, spectrum, eye,
    magnetics, inverter, mask zone, logic decoder and signal simulator lines kept the values read before the bad one
    (First Steps).
  • At 0.5 and 1 ms/div a slow edge, such as a sine of 50 to 250 Hz or mains hum picked up by a probe, put only one
    record every few seconds on the screen; the others were left out as if they had not been taken at the trigger point.
    They reach the picture again, some 30 a second.
  • Display: a Peak detect picture below the full sample rate at the short time bases (for example 1 channel at 625 MSa/s)
    is drawn from its minimum and maximum pairs, not interpolated as if they were plain samples.
  • With only noise on the source, the spectrum's THD, SNR, SINAD, SFDR and ENOB were figures of a "fundamental" found
    in the noise; they now show no value, and :SPECtrum:ITEM? answers 9.9E37
    (SCPI and VXI-11).
  • An SPI decoder on a record with too few samples for its clock no longer lists words that are none. At 200 kbit/s in a
    record of 125 kSa/s a frame of 16 words read "05 05 05 05" and the frames of one word around it gave nothing. Where
    more chip-select frames have clock edges and no whole word than frames with a word, the band now says "Frames shorter
    than a word: too few samples for the clock, or Data bits", and so it does for Data bits above the frame's length.
  • :TIMebase:SCALe from a script or the MCP tool keeps the point of Horizontal > Expand around in place and the zoom
    inside the new time base, as the knob and the number pad do; before, the delay stayed as it was. Found by W4R's OOS
    Plus fork (SCPI and VXI-11).
  • Trigger: on a slow, noisy signal (mains hum picked up by a probe, a 50 Hz sine at 0.5 ms/div) a record the hardware
    triggered on the other slope no longer shows as a mirrored trace among the others (#581, #578).
  • A new volts per division, offset or channel while the scope waits for a trigger draws the last record again at its
    own volts. The picture kept its old height and place before, so the trace read wrong values against the grid until
    the next trigger (First Steps).
  • A new time base while the scope waits for a trigger draws the last record at its own times, in the middle of the
    wider screen, instead of stretching it over the whole screen (First Steps).

Built by CI from the tag, on arm64, signed with the community key of the repository.
Still to add before this draft is published, the sources a release carries:

linux-6.12.115-base.tar.xz
busybox-1.37.0.tar.bz2
dropbear-2026.94.tar.bz2
openssh-10.6p2.tar.gz
libmnl-1.0.5.tar.bz2
libnftnl-1.2.9.tar.xz
nftables-1.1.3.tar.xz
kexec-tools-2.0.32.tar.xz
e2fsprogs-1.47.2.tar.xz
wpa_supplicant-2.11.tar.gz
mesa_26.0.8.orig.tar.*|mesa-26.0.8.tar.*
dbus-1.16.2.tar.xz
bluez-5.82.tar.xz
expat-2.11.0.tar.xz
glib-2.84.4.tar.xz
tests-machine-id.patch
flaky-tests.patch
pcre2-10.46.tar.bz2
gettext-0.24.1.tar.xz
cwd.patch
skip-tests-musl.patch
skip-meaningless-pthread-rwlock-test.patch
libnl-3.11.0.tar.gz
libnl3-musl.patch
glibc_*.dsc
glibc_*.orig.tar.*
glibc_*.debian.tar.*
gcc-[0-9]*_*.dsc
gcc-[0-9]*_*.orig.tar.*
gcc-[0-9]*_*.debian.tar.*
libdrm_*.dsc
libdrm_*.orig.tar.*
libdrm_*.debian.tar.*
zlib_*.dsc
zlib_*.orig.tar.*
zlib_*.debian.tar.*
SHA256SUMS

Don't miss a new oos release

NewReleases is sending notifications on new releases.