github dekajoule/oos mho-dev-52
mho-dev-52: four channels in step (MHO900 test build)

latest release: mho-dev-53
pre-release46 minutes ago

Superseded by mho-dev-53.

Development build for testing on an MHO900. Not a release. Built from the branch oos-mho: the fixes from hp197's tests of mho-dev-35 and -37 and his pull requests #561 and #562 (issue #4). Please test carefully and report back with oos-mho-report (or mho-setup.py <scope-ip> report).

Verified on hp197's MHO98 (mho-dev-35): fan, counter, peak detect's picture, 50 and 100 ms/div with the edge at the trigger point, roll, the generator board found, generator 1's offset.

Fixed in this build, not yet verified on hardware:

  • Four channels in step (found by JimmyPedersen): the read-out delay words now go to every channel's lane (OOS wrote them to one only). On his MHO934 CH3 lay 6 to 7 samples behind the others with all four on. A file /oos/mho-delay-one-lane brings the old behaviour back for a comparison.
  • Cursors: - and + on a voltage cursor now move it the right way (+ raises the value; found by JimmyPedersen).
  • Counter: it counts extra edges when a flat part of the signal lies exactly on the screen's centre line (its comparator's threshold); move the trace a little and it counts right. Not fixed yet.
  • Generator 2 (G2) is driven like generator 1, both at once (by hp197, #572; verified on his MHO98).
  • Generator offset: changing a channel's setting no longer resets generator 1's offset to 0 V (found by hp197).
  • Sweep: Auto and Normal reach the board (dev-48).
  • The two ADCs aligned (found by hp197): the alignment word now goes to both ADCs (OOS wrote it to one only). On his MHO98 that ended the frames triggered on the wrong edge. A file /oos/mho-adc-align (hex) tries another value; the best one differs from start to start, and measuring it at every start as the stock system does is still to come.
  • The SCU mode word 0x4008 carries the record's interleave factor (dev-47).
  • 50 and 100 ms/div: the recording took them in roll mode; OOS now makes them ordinary records with the trigger in the middle. Verified on hp197's MHO98: the 0 V stretch is gone.
  • Logic analyser: its sample rate is now set (a quarter of the analog rate, measured by hp197 under the stock system), and its words get the matching time axis. The log still shows what the LA words hold every 50 frames.
  • The empty start of some frames at 50 to 200 µs/div (Normal and peak detect): frames where the board triggered away from the trigger point are no longer shown. The log counts them, to find the cause.
  • 50 ms/div: the 0 V stretch after the trigger point is not fixed yet; the log now says where the record holds one word for long.
  • The generator's shape DC no longer crashes the screen.
  • Peak detect: the measurements no longer count twice the frequency.
  • Memory depth: a fixed depth takes the nearest rate the unit's recording has (it stayed at 10 k before).
  • Generator 1 puts out every shape: sine, square, ramp (with its symmetry), pulse (with its duty), arbitrary, noise and DC, also after a start by OOS (the output's scale words are set now, #562). Verified on hp197's MHO98: sine, square, ramp, pulse, noise and DC at the right amplitude and frequency. Up to 2 Vpp, ±1 V offset.
  • Bode with the built-in generator: verified on hp197's MHO98 through a cable (flat within +0.14/−0.05 dB from 100 Hz to 1 MHz).

Still open: the logic analyser on a unit (not yet tried with this build), 50 Ω (set it on the screen: OOS's SCPI has no :CHANnel<n>:IMPedance yet), extra counts of the counter on an MHO934.

Everything else as in mho-dev-36. Install as before; the recordings stay in /oos.

File What
oos-mho900-dev-51.img the OOS image, for the install on a second card (boot/card-install.py, then boot/switch.py oos)
oos-mho900-dev-51.oosupd the same image as a signed package, for the install over ADB (oos-update gel-install)
oos-update-mho-dev-52 the install tool of this build, for the scope (aarch64)
mho-setup.py, mho-poll the set-up in three steps: record under the stock system, install under OOS, report afterwards (docs/mho.md "Setting a unit up")
mho-check.py, mho-settings.py the recording tools mho-setup.py record runs
MHO900_OOS_Install.GEL the stick install. Not known whether the stock MHO900 update accepts it. Use the ADB or card route
SHA256SUMS checksums

Built from tag mho-dev-52 (build 20261124-205350) with kernel/build-base.sh (Linux 6.12.115; the kernel of mho-dev-35) and OOS_BOARD=mho boot/build-dev-image.sh. Sources: this tag, and the third-party sources of v0.1.14.

Don't miss a new oos release

NewReleases is sending notifications on new releases.