Development build for testing on an MHO900. Not a release. Not yet verified on hardware.
With mho-dev-23 on skluehspies' MHO98 (issue #491) 1 GSa/s and above measure, and the board's trigger fires; at 200 us/div and slower the edge is at the trigger point, from 500 MSa/s on mostly not. This build:
- reports in the screen's log, every 50 frames, how far the edge of the set slope and level nearest the trigger point lies from it (
grep 'oos-ui mho' /tmp/oos-ui.log); - puts the screen's trigger point on that edge (a software trigger on the board's frames, until the board's offset is known).
Install as before; the snapshot and the settings recording stay in /oos; oos-mho-live.
| File | What |
|---|---|
oos-mho900-dev-24.img
| the OOS image, for the install on a second card (boot/card-install.py, then boot/switch.py oos)
|
oos-mho900-dev-24.oosupd
| the same image as a signed package, for the install over ADB (oos-update gel-install)
|
oos-update-mho-dev-24
| the install tool of this build, for the scope (aarch64) |
mho-check.py, mho-settings.py
| the recording tools (unchanged) |
MHO900_OOS_Install.GEL
| the stick install. Not known whether the stock MHO900 update accepts it. Use the ADB or card route |
SHA256SUMS
| checksums |
Built from tag mho-dev-24 (build 20261123-153403) with kernel/build-base.sh (the kernel of mho-dev-10) and OOS_BOARD=mho boot/build-dev-image.sh, Linux 6.12.115. Sources: this tag, and the third-party sources of v0.1.14.